Product description
12-inch color TFT display, up to 12 wave-forms
One level menu and easy to use
Parameters includes ECG/RESP/SpO2/NIBP/temperature/pulse rate/IBP (optional)
Maximum 720H graphic and tabular trends of all parameters
Built-in recorder and battery
Suitable for adult/pediatric/neonatal patient
Arrhythmia analysis, ST calculation and pace analysis
Anti-ESU/-defibrillator
Standard network interface, can connect to central monitor system
Optional: recorder, dual IBP, CO2 and ICG
Specifications:
- ECG
- Lead modes: I/II/III/AVR/AVL/AVF/V
- Gain: X0.25/X0.5/X1/X2
- Heart rates:
- Adult range: 15 to 300bpm
- Neonatal range: 15 to 350bpm
- Resolution: 1bpm
- Accuracy: ±1%
- ST measurement range: -2.0, ±2.0mV
- Accuracy: ±10%
- Pacemaker: supported
- Sweep speed: 12.5/25/50mm/second
- Bandwidth:
- Diagnostic range: 0.05 to 130Hz
- Monitor range: 0.5 to 40Hz
- Surgery range: 1 to 20Hz
- SPO2
- Measure range: 0 to 100%
- Resolution: 1%
- Accuracy: 2% (70 to 100%)
- Pulse rate range: 20 to 250bpm
- Resolution: 1bpm
- Accuracy: ±3bpm
- RESP:
- Measure range: 0 to 150brpm
- Resolution: 1brpm
- Accuracy: ±2%
- Temperature:
- Measure range: 0 t o50°C
- Accuracy: ±0.1°C
- NIBP:
- Work mode: manual, automatic and continuous
- Unit: mmHg and kPa
- Measure range:
- Adult:
- SYS range: 40 to 280mmHg
- DIA range: 10 to 225mmHg
- Mean range: 20 to 240mmHg
- Pediatric:
- SYS range: 40 to 220mmHg
- DIA range: 10 to 160mmHg
- Mean range: 20 to 170mmHg
- Neonatal:
- SYS range: 40 to 135mmHg
- DIA range: 10 to 100mmHg
- Mean range: 20 to 110mmHg
- Resolution: 1mmHg
Standard packing list:
- 5-lead ECG cable
- Adult finger Spo2 sensor
- NIBP extension tube
- TEMP sensor
- Power cord
- ECG electrodes
Operating instruction:
- Selectable accessory: (option of one-time use)
- Child finger SPO2
- Child NIBP cuff
- Neonate NIBP cuff
- HS code: 90181930
Packaging:
- Net weight: <5kg
- Gross weight: <8kg
- Dimensions: 320x160x270mm
ECG
- Lead modes: I/II/III/AVR/AVL/AVF/V
- Gain: X0.25/X0.5/X1/X2
- Heart rates:
- Adult range: 15 to 300bpm
- Neonatal range: 15 to 350bpm
- Resolution: 1bpm
- Accuracy: ±1%
- ST measurement range: -2.0, ±2.0mV
- Accuracy: ±10%
- Pacemaker: supported
- Sweep speed: 12.5/25/50mm/second
Bandwidth:
- Diagnostic range: 0.05 to 130Hz
- Monitor range: 0.5 to 40Hz
- Surgery range: 1 to 20Hz
SPO2
- Measure range: 0 to 100%
- Resolution: 1%
- Accuracy: 2% (70 to 100%)
- Pulse rate range: 20 to 250bpm
- Resolution: 1bpm
- Accuracy: ±3bpm
RESP:
- Measure range: 0 to 150brpm
- Resolution: 1brpm
- Accuracy: ±2%
Temperature:
- Measure range: 0 t o50°C
- Accuracy: ±0.1°C
NIBP:
- Work mode: manual, automatic and continuous
- Unit: mmHg and kPa
- Measure range:
Adult:
- SYS range: 40 to 280mmHg
- DIA range: 10 to 225mmHg
- Mean range: 20 to 240mmHg
Pediatric:
- SYS range: 40 to 220mmHg
- DIA range: 10 to 160mmHg
- Mean range: 20 to 170mmHg
Neonatal:
- SYS range: 40 to 135mmHg
- DIA range: 10 to 100mmHg
- Mean range: 20 to 110mmHg
Resolution: 1mmHg
Lead modes: I/II/III/AVR/AVL/AVF/V
Gain: X0.25/X0.5/X1/X2
Heart rates:
- Adult range: 15 to 300bpm
- Neonatal range: 15 to 350bpm
Resolution: 1bpm
Accuracy: ±1%
ST measurement range: -2.0, ±2.0mV
Accuracy: ±10%
Pacemaker: supported
Sweep speed: 12.5/25/50mm/second
Adult range: 15 to 300bpm
Neonatal range: 15 to 350bpm
Diagnostic range: 0.05 to 130Hz
Monitor range: 0.5 to 40Hz
Surgery range: 1 to 20Hz
Measure range: 0 to 100%
Resolution: 1%
Accuracy: 2% (70 to 100%)
Pulse rate range: 20 to 250bpm
Resolution: 1bpm
Accuracy: ±3bpm
Measure range: 0 to 150brpm
Resolution: 1brpm
Accuracy: ±2%
Measure range: 0 t o50°C
Accuracy: ±0.1°C
Work mode: manual, automatic and continuous
Unit: mmHg and kPa
Measure range:
SYS range: 40 to 280mmHg
DIA range: 10 to 225mmHg
Mean range: 20 to 240mmHg
SYS range: 40 to 220mmHg
DIA range: 10 to 160mmHg
Mean range: 20 to 170mmHg
SYS range: 40 to 135mmHg
DIA range: 10 to 100mmHg
Mean range: 20 to 110mmHg
Accuracy: ±5mmHg
5-lead ECG cable
Adult finger Spo2 sensor
NIBP extension tube
TEMP sensor
Power cord
ECG electrodes
Selectable accessory: (option of one-time use)
Child finger SPO2
Child NIBP cuff
Neonate NIBP cuff
HS code: 90181930
Net weight: <5kg
Gross weight: <8kg
Dimensions: 320x160x270mm