HV-300A (LCD) Electrosurgical Generator HV-300A (LCD) Electrosurgical Generator
Versatility in office-based Electrosurgery
For general electrosurgery needs in the office, clinics, or emergency room, the HV-300A generator provides exceptional performance, reliability, and safety benefits in a compact, easy to use unit.
EaseEase and quick to set up; compact and light
Save time with automatic access to previous used power settings and modes
Provides precise surgical effect with power control dials
FlexibilityPerforms most common office-based electrosurgical procedures
Includes both monopolar and bipolar outputs
ReliabilityManufactured to the highest standards of quality in electrosurgical technology
Ensure consistent clinical performance over a variety of procedures
SafetyIncreased patient safety with a return electrode contact quality monitoring system (REM). The REM system continually monitors patient impedance levels and deactivates the generator if a fault in the patient/return electrode contact is detected.
High Frequency leakageBipolar: less than 60mA
Monopolar: less than 150mA
Low frequency leakageSource current, patient leads, all outputs tied together
Normal polarity, intact chassis ground
<10uA
Normal polarity, ground open
<50uA
Reverse polarity, ground open
<50uA
Sink current, 140V applied, all inputs
<50uA
Power readoutsAgree with actual power into rated to within ±20% or 5W, whichever is greater, Dossage Error monitor.
Unique features of the HV-300A Electrosurgical GeneratorImproved performance at lower power settings minimizes the risks of tissue damage and neuromuscular stimulation
Easy to use, cost effective system when performance is required and price is of equal concern
Four cut modes offer surgeons a variety of choices:
Pure cut for a clean, precise cut
Blend 1, 2, 3 for cutting with hemostasis
Two Coag modes:
Fulgurate for lower voltage coagulation requirements
Spray for coagulating large tissue areas with superficial of necrosis
Two bipolar modes:
Bipolar cut and coag: with the active electrode formed one confined circuit to perform the surgery; minimized damage to tissue by limiting the amount of tissue involved in the electrosurgical circuit.
Versatile system that is uniquely compatible with other, including:
Under water cut socket (ENDO cut)
HV argon congulation system
Optimumm Smoke Evacuation System-Clean air in the operation theatre
Technical SpecificationsOutput Frequency 540Khz
Line Voltage: 110V/220V~
Line Frequency 50/60Kz
Weight and DimensionsHeight: 14cm
Width: 32cm
Length: 41cm
Weight:
<8.5kgs
Applicable rangeApplicable for general surgery, thoracic surgery, orthopedics, obstetrics and gynecology, urologic surgery, cerebral surgery, microsurgery, ENT surgery and etc.
Output Characteristics
|
Mode |
Maximum P-P Voltage |
Rated Load (OHMS) |
Maximum Power (Watts) |
Crest Factor (Typical) |
Bipolar |
Standard Coag |
550 |
100 |
70 |
1.6 |
Cut |
700 |
100 |
70 |
1.6 |
Monopolar Cut |
Pure |
1900 |
300 |
300 |
1.6 |
Blend 1 |
2300 |
300 |
250 |
2.5 |
Blend 2 |
2500 |
300 |
200 |
2.8 |
Blend 3 |
2800 |
300 |
150 |
2.9 |
Monopolar Coag |
Fulgurat |
2800 |
300 |
120 |
2.9 |
Spray |
7000 |
300 |
100 |
6.2 |