Uv Lamp Test Chamber , UV Aging Test Chamber
1. Product description
It is the most widely used UV Weathering tester in the world, there are thousands of ultraviolet aging testers coming into use. It has become the standard of the UV test, fit standards of
ISO, ASTM,DIN,JIS,SAE,BS,ANSI,GM,U.S.GOVT,ect. It can simulate the damage of sunshine, rain, dew to the material quickly and exactly. Damage types including bleaching, discoloration, loss of light, powder, crack, turbidity, bubble-based, embitterment, strength decline and oxidation need to be engender in months or years but can engender in a few days or weeks.
Stainless Steel UV Aging Test Chamber with Balance Temperature Humidity Control
2. Technical parameters:
Accelerated Weather UV test machine
Studio Size
|
400W×1140D×390Hmm
|
Outside dimension
|
500W×1300D×1360Hmm
|
Materials
|
Inside and Outside SUS#304 stainless steel
|
Temperature range
|
RT+10~70°C
|
Temperature uniformity
|
±2°C
|
Temperature fluctuation
|
±0.5°C
|
Temperature control
|
PID SSR control
|
Humidity range:
|
≥90%RH
|
Controller
|
Korean TEMI 880 programmable controller
|
Control mode
|
Balance temperature humidity control (BTHC)
|
Test cycle setting
|
Illumination, condensation and water spray test cycle is programmable
|
Lamp power
|
40W/Piece
|
Distance from sample to lamp
|
50±2mm (adjustable)
|
Centre distance between the lamp
|
70mm
|
Irradiance
|
1.0W/cm2
|
Wavelength ultraviolet light
|
UV-A: 315-400nm; UV-B: 280-315nm (8pcs, 1600h lifetime)
|
Effective irradiation areas
|
640×1140mm
|
Irradiation black panel temperature
|
50°C~70°C
|
Standard Specimen Size
|
75×150mm or 75×3000mm (special specitications to be
described in the contact)
|
Required water depth for water channel
|
25mm, automatically control
|
Testing time
|
0~999H, adjustable
|
Power
|
220V/50Hz /±10% 5KW
|
Protection system
|
Overload short circuit protection, over temperature
protection, water lackin
|
3. Chamber Structure:
Accelerated Weather UV test machine
The inner bladder of the chamber is made of imported high-class stainless steel plate while the outer bladder is made of SUS 304 steel plate.
The heating mode adopts inner bladder trough type heating with its rapid temperature rise and even temperature distribution.
4.Chamber Features:
1).The ultraviolet aging tester is designed according to uses’ operation , it’s easy to operate, safe and reliable.
2). The thickness of specimen installation is adjustable and the specimen installation is fast and convenient.
3). The door rotating up does not interfere with the operation and the tester takes a very small space only.
4). Its unique condensation system can be contented by tap water.
5). The heater is under the container rather than in the water, which is long life, easy to maintain.
6). The water level controller is outside the box, easy to monitor.
7). The machine has truckles, convenient to move.
8). Computer programming is convenient, automatically alarming when incorrect operated or faulted.
9). It has irradiance calibrator to extend the life of the lamp tube (more than 1600h).
10). It has Chinese and English instruction book, convenient to consult.
11). Divided into three types: common, light irradiance controlling, spraying
5. Test Specimen Mounting:
The standard test specimen holder accommodates 2 panels 75×150mm. The test panels are actual the side wall of the chamber. To properly seal the chamber, it is important to have every panel holder filled with specimens or blank panels and to have panel holders in place.
6. Design criteria:
Accelerated Weather UV test machine
(1) ASTM D4587: Standard Practice for Fluorescent UV-Condensation Exposures of Paint and Related Coatings
(2) ASTM D4799: Standard Practice for Accelerated Weathering Test Conditions and Procedures for Bituminous Materials (Fluorescent UV, Water Spray, and Condensation Method)
(3) ISO 4892-1: Plastics – Methods of Exposure to Laboratory Light Sources – Part 1: General guidance
(4) ISO 4892-3: Plastics - Methods of Exposure to Laboratory Light Sources - Part 3: Fluorescent UV Lamps
(5) ASTM D4329: Standard Practice for Fluorescent UV Exposure of Plastics
(6) AATCC TM186: Weather Resistance: UV Light and Moisture Exposure
(7) ASTM D5208: Standard Practice for Fluorescent Ultraviolet (UV) Exposure of Photodegradable Plastics
(8) IEC 61345: UV Test for Photovaltalic (PV) Modules
7. Poduct image:
Product structure:
Specimen holder:
UV Lamps:
Inner structure:
our factory
|