Product description
With the most powerful functions and the most convenient operation in the market. It is equipped with the dedicated laboratory ppb sodium ion measurement electrode and reference electrode
The performance of the static measurement in the beaker has been greatly improved, compared with the traditional sodium meter; as long as the user pays attention to the electrode cleaning in use, the satisfactory result can be achieved
It can be widely used for the continuous monitoring of sodium ions in the solutions of the thermal power plants, chemical fertilizer, metallurgy, environmental protection, pharmaceutical, biochemical, foodstuff and running water
Liquid crystal display, English menu and English notepad
Computer-based: It adopts high-performance CPU chip, high-precision AD conversion technology and SMT
technology to complete multi-parameter measurement, temperature compensation and instrument self-checking
It is also featured by high precision and good repeatability
High Sensitivity: In the static water sample, it can measure 0.2ppb (8pNa); With a long stablization time, it has very stable readings
Electrode diagnosis: With a variety of calibration modes and E0 and S queries, it conducts the monitoring of electrode performance
Network function: RS232 bi-directional communication, which can be equipped with a printer or TLab8.0
Measurement range:
- 0.2 ug/L~23000mg/L
- Resolution: 0.1 ug/L; 0.01mg/L
- 0.00pNa~8.00 pNa
- Resolution: 0.01pNa 0-60℃
- Resolution: 0.1℃
- Accuracy: ±0 2 mV, ±0 3℃
- Temperature compensation range: automatic/manual: 0-99.9 ℃
- Repeatability: Reading ± 2.5%
- Stability time: Reading ± 2.5% /24h
- Detection lower limit: 0.2 ug/L (8pNa)
- Input impedance: >1012Ω
- Water samples: 0-60℃
- Communication interface: RS232
- Power supply: AC 220V±22V, 50Hz ±1Hz
- Working conditions:
- Ambient temperature: 0-60℃
- Relative humidity ≤85%
- Power-down memory: >5 years
- Response time: 32 seconds (25℃), 50 seconds (60℃)
0.2 ug/L~23000mg/L
Resolution: 0.1 ug/L; 0.01mg/L
0.00pNa~8.00 pNa
Resolution: 0.01pNa 0-60℃
Resolution: 0.1℃
Accuracy: ±0 2 mV, ±0 3℃
Temperature compensation range: automatic/manual: 0-99.9 ℃
Repeatability: Reading ± 2.5%
Stability time: Reading ± 2.5% /24h
Detection lower limit: 0.2 ug/L (8pNa)
Input impedance: >1012Ω
Water samples: 0-60℃
Communication interface: RS232
Power supply: AC 220V±22V, 50Hz ±1Hz
Working conditions:
- Ambient temperature: 0-60℃
- Relative humidity ≤85%
Power-down memory: >5 years
Response time: 32 seconds (25℃), 50 seconds (60℃)
Ambient temperature: 0-60℃
Relative humidity ≤85%