Product Description PCY-III-X High-Temperature Horizontal Thermal Expansion Apparatus The equipment adopts computer and intelligent instrument (single chip) techniques to achieve detect and control the overall process of physical quantity displacement and temperature experiment. Moreover, it can realize off-line working, online realize automatic testing. Work in the environment of English Windows xp/sp2, with friendly English users operating interface; off-line state detected by intelligent instrument, manual testing. The equipment is mainly used in testing thermal expansion of ceramics, metal, alloy, graphite, refractory material, precision mold shell, core material, and other solid materials. Testing metallic material can get material phase change curve. With characteristics of full-automatic, easy to test, easy to operate, experimental data analysis, data printout, experimental data on file, data pool enrich, easy to modify and search, and so on, it provides modern test method for testing material products’ physical properties testing, scientific research, and teaching in factories and research institutions. The instrument meets the standard in GB/T3810.8-2006 /ISO10545-8: 1994 Ceramic tiles test method, part 8: Linear thermal expansion determination test requirements. Technical specifications:
1. | Single push rod dilatometer for measurement of length changes (expansion and/or shrinkage) along with operating and data acquisition software, hardware and calibration standards. |
2. | Temperature range | According to user can customize ambient ~ 1000°C/1400°C/1600°C/1700°C |
3. | Sample diameter | 6~8mm |
4. | Sample length | 50mm(max) or 25mm |
5. | Atmosphere | According to the temperature you order can be vacuum, Static and dynamic; Inert or reaction gas (not for below 1000°C) |
6. | Measuring range and error | ± 5 mm ± 0.1%, ± 2.5 mm ± 0.1% |
7. | Heating element | According to user can be resistance wire up to 1000°C, Silicon carbide up to 1400°C, or MoSi2 up to 1700°C. |
8. | Sample holder and probe rod | Quartz or Al2O3 according to the temperature you order |
9. | Dilatometer orientation | Horizontal |
10. | Expansion/shrinkage value measuring resolution | 1um/digit |
11. | Thermocouple | Type B |
12. | Vacuum level | According to user can be 10-1mbar or better |
13 | Water cooling | Closed loop water cooling unit |
14 | Heating rate | 1 ~ 30 oC /minute, accuracy±1 oC |
15 | Calibration standards | Quartz/Corundum standard sample |
16 | Temperature control | Experimental process controls parameters, including PID parameters and the way of communication. Arbitrary adjust, set, lock are realized by intelligent instrument in the whole process. Modifying the parameters is achieved by the host computer (the computer). |
17 | Data Acquisition | Trough onboard computer with printer |
18 | Power supply | 220V, 50/60HZ |
It is for precise testing materials expansion or shrinkage in the process of heat treatment. It is suitable for the testing of ceramic, building materials, glass, refractories, metal or alloy, etc. By applying PCY-Ⅲ, you can accomplish the works as follows: thermal expansion coefficient, softening temperature, the sintering kinetics research, glass transitional temperature, phase changes, density changes, and sintering rate control (RCS).
Packaging & Shipping 1 Delivery date: 20days 2.Payment terms: TT 30% in advance as the deposit, the balance to be paid before shipment. 3.Package: standard exporting plywood case.
We are specialized in tester, instrument&equipment of ceramics, inorganic non-metallic materials, glass, cement&engineering testing.