1 / 7
Model No. : | 752N Plus |
---|---|
Brand Name : | NANBEI |
Support customization : | Oem |
Zhengzhou, Henan, China
Product description
Laboratory Spectrometer 752N Plus UV VIS Spectrophotometer
Introduction
752N UV-VIS Spectrophotometer been widely used in colleges and enterprises for general quantitative analysis and experiments based in absorbance measurements.
● High quality silicon photometric diode detector and 1200 lines/mm diffraction grating ensure the high quality accuracy and precision.
● Backlit LCD display for an easy readout.
● Automatic 0A and 100%T.
● Easy switching of transmittance, absorbance and concentration modes, just by pressing one key.
● Direct concentration read-out and concentration factor setting function.
● Large sample compartment, for 5 - 100mm path length cuvettes with optional holders.
● Easy to change the halogen lamp or deuterium lamp by the user himself.
Main Features
● Sample compartment for 5-50mm cuvettes
● Precise automatic T/A changeover
● Automatic zero and full scale adjustment
● Equipped with USB port
● Direct concentration read-out and concentration factor setting function
● Automatic light gate technology to protect photoelectric sensors
7-inch multi color touch-screen
752N Plus UV-VIS Spectrophotometer has the holographic blazed grating monochromator has the advantages of high wavelength accuracy, good monochromaticity and low stray light.
● 7-inch multi color touch-screen, good human-computer interface.
● Linear regression method and coefficient method are added to the concentration test method.
● USB interface is added, and large capacity memory can store 30 concentration curves.
Adopt microcomputer measurement system
● Adopt microcomputer measurement system, with high conversion accuracy of T-A, automatic adjustment of
● 0% T and 100% T, concentration factor setting and concentration direct reading.
● High accuracy, Good Reproducibility and Stability of Measurement Readings.
● Automatic light gate technology, No need blackbody, to protect the photoelectric sensor
Specification
Photometry:
|
Split Beam
|
Monochromator Type:
|
Czerny-Turner
|
Focal Length:
|
160mm
|
Grating:
|
1200 lines/mm
|
Detector:
|
Silicon Photocell
|
Spectrum Bandwidth:
|
2nm
|
Wavelength Setting:
|
Manual Turn Knob
|
Wavelength Range:
|
200-1000nm
|
Wavelength Accuracy:
|
±2nm
|
Wavelength Repeatability:
|
≤1nm
|
Stray Light :
|
≤0.1%T( at 220nm NaI, 360nm NaNo2 )
|
Photometric Range:
|
0.0-200.0% T
|
-0.097-4.000 A
|
|
0.000-9999 C
|
|
Photometric Accuracy:
|
±0.5% T
|
±0.004 Abs (0-0.5A)
|
|
±0.008 Abs (0.5-1A)
|
|
Photometric Repeatability:
|
≤0.2% T
|
0.002 Abs (0-0.5A)
|
|
0.004 Abs (0.5-1A)
|
|
Noise:
|
100%(T)noise≤0.3%(T) , 0%(T)noise≤0.2%(T)
|
Power:
|
AC220V±22V 50Hz±1Hz, 120W
|
Instrument Size& Weight
|
580mm X 460mm X 345mm 0.1M3 16.5kg
|
|
|
|
Zhengzhou, Henan, China
Send your inquiry to this supplier