Product description
Features:
Compatible with syringes of most of the standard brands
Automatic calibration for syringes
Three working modes: rate control mode, time control mode anddose weight control mode (ug/kg/min or mg/kg/h)
Built-in rechargeable battery (lithium-ion polymerbattery)
Alarm functions: occlusion, empty, near empty, low battery,infusion over, syringe loose and more
Three levels of the occlusion limited pressure: high, middleand low
Three levels of alarm sound: high, middle and low
Syringe pump goes into KVO mode automatically as infusion iscompleted
Purge/bolus function
Technical specifications:
Flow range (rate mode):
- 50mL syringe: 0.1-1800.0mL/hr (0.1mL/hr step)
- 20mL syringe: 0.1-800.0mL/hr (0.1mL/hr step)
- 10mL syringe: 0.1-400.0mL/hr (0.1mL/hr step)
Accuracy: ± 2% (after correct calibration)
Time setting mode range:
Volume: 0.1-199.9mL, 0.1mL step
Time: 1-1999min, 1min step
50mL syringe: 0.1-1800.0mL/hr (0.1mL/hr step)
20mL syringe: 0.1-800.0mL/hr (0.1mL/hr step)
10mL syringe: 0.1-400.0mL/hr (0.1mL/hr step)
Weight setting mode range:
Sol. volume: 0.1-199.9mL, 0.1mL step
Drug mass: 0.1-999.9mL, 0.1mL step
Dose: 0.01-99.99ug/kg/min, 0.01ug/kg/min step,0.01-99.99mg/kg/h, 0.01mg/kg/h step
Weight: 0.1-99.9kg, 0.1kg step 100-300kg, 1kg step
Purge rate: 1,800mL/h (50mL), 800mL/h (20mL), 400mL/h(10mL)
Volume limits range: 0-1999.9mL, 0.1mL step
Bolus volume: 1.0-100mL, 0.1mL step
Power supply: AC 100-240V, 50/60Hz
Internal lithium-ion polymer battery: DC 11.1V, e1,600mAh
Battery backup is approximately 8 hours (when using a fullycharged battery and running at 5mL/h)
Power consumption: d25VA
Size: 280 x 160mm