Model No. : | C5 |
---|---|
Brand Name : | Landcom |
Product description
Item | Specification |
Magnet | |
Magnet Field strength | 0.35 T |
Homogeneity | <20 ppm(peak-peak) in 40X35cm DSV |
5GS line | < 3m |
Patient Gap(Cover to cover) | 40 cm in height |
Operation software ( Supernova R1) | |
Acquisition Techniques | Flow Compensation Turbo Multislice Presaturation Dual band preset Cardiac Gating Respiratory Gating Half scan Partial echo Rect FOV Rect Scan Matrix Multi-Scan Dynamic Imaging MTC Phase Order (Sequence/Central Out) Three dimension pilot Automatic field adjustment Reverse slices acquisition & display |
Image display | Adjustment on window W/L Scroll Alignment Pan Zooming Rotation Multi-format of display (1 HOR-2,VER-2,4,9,16) Annotation Statistics (ROI) Cine display Pilot image display in PIP mode Show position line on any in-study images Different serials compare |
Image processing | Smart Background Prune MIP(Maximum Intensity Projection) Full set of image analysis tools Realtime zoom Selection of image filters ROI analysis tools Batch mode processing Multi-Page film composer |
DICOM 3.0 | Storage, Worklist Management, Print |
Protocol package | |
Spin echo 2D/3D |
|
Gradient Echo(GE 2D,GE3D) |
|
Fast Spin Echo (FSE) |
|
Inversion recovery(IR-SE ,STIR) |
|
Fast Inversion Recovery (IR-FSE,FLAIR) |
|
MR Angiography(TOF 3D,2D) |
|
DWI |
|
Echo Plane Imaging (EPI) Single-Shot FSE(MRCP,MRU,MRM) |
|
General Image Parameters | |
Scan orientation | Transverse Sagittal Coronal Oblique |
Maximum Acquisition Matrix | 1024X1024 |
Minimum TE (ms) | 2.8 |
Minimum TR (ms) | 5.6 |
Maximum Slice Number | 256 |
FOV | 5mm-400mm |
Minimum Slice Thickness (2D) | 1mm |
Minimum Slice Thickness (3D) | 0.1m |
Conditions | |
Electrical Requirements | |
Input voltages | 415/380VAC±10% @ 50 Hz, 3-phase, without N 480VAC±10% @ 50 Hz, 3-phase, without N |
System wiring requirements | 3-phase, 4 wires (3 power + 1 isolated earth ground) |
Scan room | |
Room size (depth X width X height) | 6m X 5m X 3m |
Environment requirements | Air conditioning 20---26℃,non- condensing, |
Operator room | |
Room size (depth X width X height) | 3m X 5m X 3m |
Environment requirements | Air conditioning 20---26℃,non- condensing, |
Equipment room | |
Room size (depth X width X height) | 3m X 2.5m X 3m |
Environment requirements | Air conditioning 20---26℃,non- condensing, |
Send your inquiry to this supplier