DGA 2010 Oil Gas Chromatography/ Insulating Oil Gas Chromatography/ Transformer Oil Gas Chomatography
on about the DGA2010:
We appreciate your choice for DGA2010 Gas Chromatography; your kind attention is invited to this instruction manual before running the equipment.
Equipped with large LCD displayer, DGA2010 is a micro-computer controlled GC with programmed 8-grade temperature rising function. Adopted with the advanced technology, DGA2010 has the function of power off protection, over-temperature protection, [0℃" protection, non-gas protection, auto ignition, and etc.
It has the characteristics of reliable performance, compact design, convenient operation, and good expandability. It has the special function of temperature tracking for column. Its standard configuration is FID & TCD while with the expanding detectors like ECD, FPD, and NPD as well as methane transforming furnace. It is also capable of configuring FID amplifier.
Now DGA2010 has been widely used in the industries like oil, chemistry, pesticides, wine, environment protection, epidemic prevention, universities, scientific research and etc.
1. Working Environment
Temperature: 0~30℃
Relative Humidity: <85%
No Strong Electromagnetic Field
No Corrosive Gas
Stable Working Table without Vibration
Power Supply: AC220V±10%,50Hz±0.5Hz
Power Consumption: 2KW Technical Parameters
1. Temperature Controlling
Chamber Temperature of Chromatographic Column
Temperature Range
|
RT + 8℃~420℃
|
Temperature Increment
|
1℃
|
Temperature Accuracy
|
±0.1℃ when < 200℃; ±0.2℃ when > 200℃
|
Temperature Uniformity
|
<1%
|
Temperature Error between Displaying & Real Temperature
|
<2%
|
Temperature Ramp Grades
|
8 Grades Linear
|
Temperature Ramp Rate
|
1~30℃/min with increment of 1℃/min
|
Time Range
|
0~255 min with increment of 1 minute
|
Temperature Repeatability
|
<2%
|
Max Heating Power
|
1500W
|
Temperature Sensor
|
PT100
|
Temperature of Vaporizing Room
Temperature Controlling Precision
|
±0.1℃ when <200℃; ±0.2℃ when >200℃
|
Temperature Range
|
<420℃
|
Heater
|
Standing Heating Facility with 100W Heater
|
Temperature Sensor
|
PT100
|
Chamber Temperature of FID
Temperature Controlling Precision
|
±0.1℃ when <200℃; ±0.2℃ when >200℃
|
Temperature Range
|
<420℃
|
Heater
|
Standing Heating Facility with 2×90W Heaters
|
Temperature Sensor
|
PT100
|
Sensitivity
|
S≥3000mv·ml/mg(Benzene, H2)
|
Noise
|
≤0.02mv
|
Drift
|
≤0.1mv/h
|
Built-in Amplifier
Semi-pervasion 100Ω bridge tungsten-rhenium wire
Constant Current Source Supply
|
Detection Limit
|
M≤2×10-11g/s (Benzene/SO2)
|
Noise
|
≤5×10-13A
|
Drift
|
≤5×10-12A/30min
|
Detection Limit on Programmed Temperature Control Mode
|
M≤1×10-9g/s
|
Drift on Programmed Temperature Control Mode
|
≤5×10-12A
|
All-collecting Polar Corundum Nozzle
Platinum Ignition Wire
|
4. Weight & Dimension
Dimension: 660(Width)× 460(Height)× 430(Depth)
Weight: 60kg
5. Optional Facilities
Recorder
-- Bench-top two-way recorder (Min Measurement: 1mv; Stroke Time: 1s)
-- Chromatogram Data Processor
-- Chromatogram Data Working Station (need to work with PC)
Gas Source
-- Nitrogen, Hydrogen, and Air tanks with Decompression Valves
-- Nitrogen, Hydrogen, and Air Generators
6-ways Gas Injector
Methane Reforming Furnace