1 / 1
Model No. : | Infant Phototherapy Incubator 8502H |
---|---|
Brand Name : | MEDECO |
Feature : | Reusable |
Product description
Medical Equipment Infant Phototherapy Incubator 8502H
1) Computer servo control box temperature, skin temperature control, humidity control
2) Separately display set temperature, real-time temperature and humidity
3) >37º C high temperature setting mode
4) Self-test function
5) Alarm and acousto-optic prompt of over-heating, deviation, circulation, sensor, power outage
6) Infinitely adjustable baby bed tilt
7) Aluminous tank
8) Integral aluminum tank can be taken out for cleansing and disinfection
9) Protecting rail
10) RS-232 joint
11) Oxygen input joint
12) Wind way humidified
13) Streamline trilby
14) Organic cover
15)Optional Blue-ray therapeutic apparatus that keep warm and provide treatment for jaundice, which equipped with timer
16) Filtrate UVA/UVB strong energy to avoid the harm to infant skin, the phototherapy effect is better than any other same product in China
17) OSRAM light tube imported from Germany is optional.
18) Double open door-ray film box, double organic cover, and emergency power are optional. 19) Equipped with double blue-ray, elevating device
INFANT CUBATOR TECHNICAL PARAMETER LIST
YXK-7G
YXK-2000G
YXK-2000GA
8502
YXK-5GB
YXK-6G
Power standard
220V/50Hz
220V/50Hz
220V/50Hz
220V/50Hz
220V/50Hz
220V/50Hz
Power Consumption
≤320VA
≤450VA
≤450VA
≤650VA
≤320VA
≤320VA
Environment Temperature
20-30º C
20-30º C
20-30º C
20-30º C
20-30º C
20-30º C
Temperature display range
25-45º C
25-45º C
25-45º C
25-45º C
25-45º C
25-45º C
Temperature control range
25-37º C
25-37º C
25-37º C
25-37º C
25-37º C
25-37º C
The control range of the skin temperature
25-37º C
25-37º C
25-37º C
25-37º C
25-37º C
25-37º C
>37º C High temperature control range
37.1-39º C
37.1-39º C
37.1-39º C
37.1-39º C
37.1-39º C
37.1-39º C
Temperature fluctuation range
<0.5 º C
<0.5 º C
<0.5 º C
<0.5 º C
<0.5 º C
<0.5º C
Bed surface temperature Error
<1.5º C
<1.5º C
<1.5º C
<1.5º C
<1.5º C
<1.5º C
Humidity display range
-
0-90%RH
0-90%RH
0-90%RH
-
-
Humidity display accuracy
-
+-10%RH
+-10%RH
+-10%RH
-
-
|
||
|
||
|
|
|
|
||
|
|
|
|
||
|
|
|
|
|
|
|
||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
Send your inquiry to this supplier