Portable X-ray testing equipment
Portable X-ray testing equipment
Portable X-ray testing equipment

1 / 1

Portable X-ray testing equipment

Get Latest Price
Send Inquiry
Model No. : XXQ/XXG/XXH series
Brand Name : HUARI

Product description

Specifications Portable testing equipment

1. The controller works without noise

2. Suitable with most brand of power generator

Portable testing equipment  Construction  X-ray head Controller accessories  Main technical parameters  Accessories

No.

Nmae

Count

1

Power cable(3×2.5 cores,   10m)

1

2

Controller and tube head Connection

(10×2.5 cores  25m)

1

3

Fuses 2A,10A,20A

5 pieces every

4

Ground cable(5m)

1

5

Power switch key

2

6

Door interlock cable plug

1

    The merit of HR-900 X-ray controller  1, Use IGBT working in high frequency instead of SCR. So the X-ray tube voltage and current keeps very stable even if the AC220 power supply voltage is violence changing in a wide range. 2, Has a higher power factor than controller with SCR. 3, Suitable with most brand of power generator 4, Has a real time clock and EEPROM in the controller that can remember operation record automaticaly. The controller can decide whether season is needed by this record.         The compare of controller with IGBT and controller with SCR

 

Ordinary controller with SCR.

HR-900 controller with IGBT

Power factor

0.23-0.45

0.66

Power factor constant

No, power factor decrease along with tube voltage decrease

Yes

power consumption

High

Low

Generator (When There is not power source in the open air)

Suitable with a few brand of power generator

Need adjust synchronization

Need large power generator, normal 5000W

Waveform of controller instability

Suitable with most brand of power generator

No need adjust synchronization

Need small power generator

Waveform of controller stability

Output tube voltage influenced by power fluctuations

Large

No influenced

Both AC and DC

No

Yes

Number of adjustment potentiometer

6

2

Limiting output current function

No

Yes

Capacity of resisting impact of fault tube

weak

Because high voltage generator primary coil can automatically limit current. So capacity of resisting impact is strong. The controller almost is not influenced by high voltage fault and X-Ray tube fault.

Automatic season according to stopping time

No

Yes

No power off until full cooling time. Force wind cooling 

No

Yes

Directly shut off high voltage by key

No

Yes

 x ray equipment,ndt equipment,x ray flaw detector
Send Inquiry

Product Alert

Subscribe to your interested keywords. We will send freely the latest and hottest products to your Inbox. Don't miss any trade information.

版权所有 © 宁波全贸信息技术有限公司 浙ICP备12012821号-1 浙B2-20200628